Original Article OPEN ACCESS
Evaluation of lithium determination in three analyzers: flame emission, flame atomic absorption spectroscopy and ion selective electrode
1Department of Biochemistry, National Reference Laboratory, Ministry of Health and Medical Education of Iran.
2Department of Statistics, Iran University of Medical School, Tehran, Iran.
Citation: Aliasgharpour M, Hagani H. Evaluation of lithium determination in three analyzers: flame emission, flame atomic absorption spectroscopy and ion selective electrode. North Am J Med Sci 2009; 1: 244-246.
Doi: 10.4297/najms.2009.5244
Availability: www.najms.org
ISSN: 1947 – 2714
Abstract
Background: Lithium carbonate salt has become an increasingly important substance in the treatment of manic depressive disorders, and its narrow safe therapeutic dose range has caused laboratories to monitor the serum concentration carefully. In the present work we compared the efficiency of three analyzers in lithium measurement. Aims: The present work was to compare the efficiency of flame emission (FES), flame atomic absorption spectroscopy (FAAS) and ion selective electrode (ISE) in lithium measurement. Materials and Methods: The three different analyzers were used. The samples assayed were Seronorm-Trace elements serum as quality control material and patients’ serum. All chemicals had a grade suitable for trace metal analysis. Results: The percent recovery of added lithium in pooled patients’ serum was higher for ISE than for FASS and FES. We also obtained a higher average lithium concentration for patients’ serum samples measured by ISE than for FAAS and FES. Paired t-test results revealed a significant difference (p< 0.001) for patient sera analyzed with FAAS and ISE. Conclusion: We report higher results for ISE than the other two analyzersandconclude that the choice between the two flame methods for patients’ serum lithium determination is arbitrary, and that FES analyzer is a more attractive routine alternative for lithium determination than FAAS because of its cost and ease of performance. In addition, the results obtained by ISE are precise. However, its accuracy may depend on other interfering factors.
Keywords: Lithium, Atomic Absorption Spectroscopy, Ion Selective Electrodes, Flame Photometry.
Correspondence to: Mehri Aliasgharpour (MS-ASCP), Department of Biochemistry, National Reference Laboratory, Ministry of Health and Medical Education of Iran, Tehran, Iran. Tel.: (+98) 218889 2070. Email: mehri9@ yahoo.com
Introduction
For the effective treatment of manic depressive disorders, administration of lithium salts was approved by the US Food and Drug Administration in 1970 and they have been used widely since then (1). Contrary to the low lithium concentration in normal sera, lithium is toxic in high concentrations and yet is ineffective for treatment if the concentration is too low (2-4). Thus, it is important to maintain the blood serum concentration at relatively narrow safe therapeutic dose range and it is apparent that accurate and precise measurement of lithium in manic patients’ sera is vital in order to assure adequate and safe treatment.
Many different methods have been introduced for serum lithium determination. Initially, flame emission spectrometry (FES) and flame atomic absorption spectrometry (FAAS) were used to determine blood serum lithium concentration (5-7). In the late 1980s, ion selective electrodes (ISEs) were developed for lithium (8), and recently a colorimetric method was developed (9). In the present work, we aim to compare and evaluate the efficiency of three analyzers, FES, FAAS and ISE, in lithium measurement.
Materials and Methods
Analyzers and Calibration of the Systems
FAAS: A Varian SpectrAA 20-Plus was used for the analysis. The instrument was calibrated with Li working standards (1, 2, 3, 4 µg/ml) made from Li stock standard solution (1g/L). Manually diluted (1:20) pool control serum samples were then aspirated into the air/acetylene flame where neutral Li atoms absorb light emitted from the Li hollow cathode lamp at 670.8 nm (6). The standard solutions and all the sera dilution were in KCL 2g /L to suppress the interfering substances.
ISE: The AVL model 9180 electrolyte analyzer was used for the analysis. A calibration step was performed for the Li configuration using the manufacturer’s suggested procedure. Afterwards, neat pool control serum samples were analyzed.
FES: The FES instrument was a CIBA-Corning 480, which was equipped with an automatic diluter. The instrument was calibrated with Multical standard (1.5 mmol/l Li concentration) solution and neat pool control serum samples were introduced to the instrument.
Other necessary chemicals and solutions used in the analysis were of a grade suitable for trace metal analysis and were purchased from Merck Co. De-ionized water (D.I-H2O) was used for the reagents preparation. The quality control serum material used for the evaluation procedures was Seronorm-Trace Elements serum (Sero AS Asker-Norway).
Precision Studies
We performed within-day and day-to-day precision studies on three analyzers by reconstituting and pooling ten Seronorm Trace Elements Serum vials—quality control material. Within-day precision was determined by analyzing a pool serum over three analytical runs (n=12) on the same working day. Day-to-day precision was determined by analyzing aliquots of pool serum (n=10) on 10 consecutive days.
Recovery Studies
For recovery studies in the three systems, we determined the concentration of the pool patients’ serum in triplicates. Then, 200 µL of Li standard solution (2.00 mmol/L) was added to the equal volume of pool serum and percent recovery was calculated.
Linearity Studies
The linearity study on each system was determined by assaying the neat pool serum 3 times, performing dilution by a factor of 2, and assaying again. Data was analyzed with linear regression using SPSS software. For the lower detection limit of each system, a sample at the lower concentration was run 20 times, and then mean and standard deviation (SD) were calculated.
Patients’ Serum Correlation Studies for Lithium
16 patient serum samples, for which Li determination had been requested, were assayed with analyzers. The samples were first analyzed with FAAS, and then underwent further analysis by other two systems on consecutive days and results were analyzed by paired t-test.
Results
With respect to the precision studies of Seronorm serum control material, the within-day precisions (CV %) were less for FAAS (1.5%) and FES (0.90%) than for ISE (1.9%). The results for between-days precision of CVs% were less for FES than for FAAS and ISE (1.3% versus 2.2% & 2.3%), Table 1. Table 2 shows the recovery of added lithium (2.0 mmol/L) for three systems. Table 3 shows the results obtained for linearity and lower limit of detection (LLD) for the analyzers. Analysis of the 16 patients’ sera is summarized in tables 4 and 5.
Table 1 Precision assay evaluation for seronorm-trace element control serum
Precision |
FAAS |
FES |
ISE |
Within day (36) Mean, mmol/L |
0.84±0.013 |
0.83 ±0.008 |
0.86±0.016 |
CV% * |
1.5 |
0.90 |
1.9 |
Day to day (10) |
|
|
|
Mean, mmol/L |
0.83±0.019 |
0.82 ±0.011 |
0.87±0.015 |
CV% * |
2.2 |
1.3 |
2.3 |
Dilution |
(1:20) + |
(1:50) ++ |
neat |
*: Obtained CV% is at analytical Li range of 0.82-0.84 mmol/L & target value of 0.83 mmol/L for seronorm-trace element control material.
+: Manual dilution; ++: Automatic dilution
Table 2 Recovery of lithium added to pool serum
Pool serum (mmol/L) Added Li (2mmol/L) |
FAAS |
FES |
ISE |
0.84±0.03 200 µl |
Measured* Recovered% |
Measured Recovered% |
Measured Recovered% |
* n=3
Table 3 Linear Regression analysis for linearity studies
|
FES |
ISE |
Slope |
0.981 |
0.982 |
LLD *(mmol/L)/Dilution |
0.061±0.001 /(1:16) |
0.095± 0.003 /(1:16) |
Dependent variable- AAS; * AAS LDL = 0.00± 0.014/ (1:32)
Table 4 Descriptive analysis for patients’ serum
Analyzers |
N |
Mean |
FAAS |
16 |
0.704± 0.26 |
FES |
16 |
0.735±0.19 |
ISE |
16 |
0.825±0.30 |
Table5 Paired samples statistics for patients’ serum lithium
Pairs |
95% CI* |
t |
DF |
p |
FAAS-FES |
-0.083 – 0.021 |
-1.283 |
15 |
0.22 |
FAAS-ISE |
-0.182 -0.061 |
-4.28 |
15 |
0.001 |
٭CI: Confidence Interval; ٭DF- Degree of freedom
Discussion
Lithium carbonate salts have become an increasingly important substance in the treatment of manic depressive disorders (1-3, 10), and its narrow safe therapeutic dose range has caused laboratories to monitor their serum concentration carefully. A variety of techniques are used to determine lithium concentration in patients’ serum samples (6, 7). The purpose of the present study was to compare and evaluate the efficiency of three analyzers, FES, FAAS and ISE, in lithium measurement.
Our obtained results indicated lower between-days percentages of CV for FES than for FAAS and ISE (1.3% versus 2.2% and 2.3%) using Seronorm serum control material. The result of within-day precision was higher for ISE than for FES and FAAS as well.
The percent recovery of added lithium for FES and FAAS was satisfactory. However, it was higher for ISE than for FASS and FES (103.4% vs. 96.2% and 94.6%).
In addition, we obtained a higher average lithium concentration for patients’ serum samples (n=16) measured by ISE than for FAAS and FES, and paired t-test results revealed a significant difference (p < 0.001) for patient sera analyzed with FAAS and ISE. In summary, our obtained results indicated higher values for ISE than the other two analyzers. These findings are in agreement with those of Serdarevic, who also observed higher results using ISE than AAS and dry-slide technology (11). We conclude that even though ISE determination is precise and easy, its accuracy may depend on other interfering factors. Furthermore, our results are in agreement with Levy and Kate (12), who indicated that the two flame methods—FES and FAAS—compared quite well and were precise.
Conclusions
We report a higher result for ISE than the other two analyzers, and summarize that the choice between the two flame methods for patients’ serum lithium determination is arbitrary. The FES analyzer is a more attractive and routine alternative than FAAS due to its cost and ease of use.