ORIGINAL ARTICLE
Year : 2014  |  Volume : 6  |  Issue : 5  |  Page : 224-228

Types and frequency of errors during different phases of testing at a clinical medical laboratory of a teaching hospital in Tehran, Iran


Department of Clinical Pathology, Central Laboratory,Imam Hospital Complex, Tehran University of Medical Sciences, Tehran, Iran

Date of Web Publication21-May-2014

Correspondence Address:
Alireza Abdollahi
Associate Professor of Pathology,Division of Clinical Pathology, Imam Hospital Complex,Tehran University of Medical Sciences, Tehran
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1947-2714.132941

Rights and Permissions
  Abstract 

Background: According to official data, 60-70% of clinical decisions about hospitalization and discharge are based on laboratory results. Aims: The objective of this study is to examine the frequency of errors before, during, and after analysis in a major medical laboratory. Materials and Methods: This descriptive, cross-sectional study was conducted throughout 2012 (January-December 2012). Errors are recorded by the Quality Control Committee in a specially designed record. Results: A total of 303,866 samples, 2,430,928 tests were received for analysis. The total number of errors was 153,148 (6.3%) (116,392 for inpatients and 36,756 for outpatients). Analysis of the results revealed that about 65.09% of the errors occur across preanalytical phase, whereas 23.2% and 11.68% are related to analytical and postanalytical phase, respectively. Conclusion: More than half of the laboratory errors are related to preanalytical phase; therefore, proper training and knowledge of intervening factors are essential for reducing errors and optimizing the quality.

Keywords: Different Pphases of Ttesting, errors, laboratory medicine


How to cite this article:
Abdollahi A, Saffar H, Saffar H. Types and frequency of errors during different phases of testing at a clinical medical laboratory of a teaching hospital in Tehran, Iran. North Am J Med Sci 2014;6:224-8

How to cite this URL:
Abdollahi A, Saffar H, Saffar H. Types and frequency of errors during different phases of testing at a clinical medical laboratory of a teaching hospital in Tehran, Iran. North Am J Med Sci [serial online] 2014 [cited 2023 May 30];6:224-8. Available from: https://www.najms.org/text.asp?2014/6/5/224/132941


  Introduction Top


Today, medical laboratories play a significant role in the healthcare system and the decision-making of clinical doctors about their patients. According to official data, 60-70% of clinical decisions about hospitalization, discharge, and prescription are based on laboratory results. [1],[2],[3],[4] Clinical tests enjoy a high status in screening, treatment follow-up, and assessment of response to treatment. [1],[2],[3],[4] Given this high percentage and the significant role, the quality of laboratory tests is of high importance. [1],[2],[3],[4] In other words, quality is the cornerstone of management in a laboratory [5],[6] and it needs to conform to the highest standards. [5],[6] Laboratory services need to be accurate, precise, and quick to prove effective. Errors might happen throughout the process. The errors are categorized into three stages depending on whether they are done before, during, or after analysis: the preanalytic, analytic and postanalytic phase errors. [6],[7]

Since there are no definite methods of sampling, most errors take place before the test, just before the sample is being prepared (it includes specimen collection, handling and processing variables, physiological variables, and endogenous variables). [6],[7],[8] Some of the preanalytical variables such as specimen collection can be controlled, while knowledge of uncontrollable variables needs to be well understood to be able to separate their effects from disease-related changes affecting the laboratory results. [7],[8]

Preanalytical errors are reported to be up to 70% in various studies. [6],[7],[8],[9] Since the quality is interconnected, precision and accuracy are not the only guarantors of the quality. From the very beginning, all three stages need to be under monitoring and quality control with precision and accuracy. The objective behind quality control is to minimize laboratory errors.

The objective of this study is to examine the frequency of errors before, during, and after analysis in a referral laboratory so that proper approaches could be developed for minimum errors.


  Materials and Methods Top


This descriptive, cross-sectional study was conducted at Imam Teaching Hospital throughout 2012 (January-December 2012). The hospital, affiliated to Tehran University of Medical Sciences, is a referral hospital for the entire country. It has a department for every medical specialty (Surgery, Internal Medicine, Pediatrics, Intensive and coronary care units, etc.). Ethics and Research Committee of the university has endorsed this study.

The sections in the hospital''s laboratory include hematology, urine biochemistry, parasitology and microbiology, general and specialized biochemistry, hormonology, serology, coagulation, and flow cytometry (FCM). Two associate professors and an assistant professor, as well as 106 employees are working in the laboratory, which provides laboratory testing for inpatients and outpatients round the clock.

The laboratory testing process starts either through a requisition from the hospital's departments or by physicians in private practice. For inpatients, the laboratory clerks process the patients' identification data including their name, age, sex, file number, and physician's name and register them in the hospital information system (HIS). If the requisition is not stat order or emergent, the following morning a trained laboratory staff collects blood samples from all hospital departments. In stat and emergent occasions, nurses or medical students collect the blood samples, which are then transported by couriers to the laboratory department. For outpatients, the physicians in private practice fill out the appropriate requisition form, which serves as a referral to the laboratory department. Then, the laboratory clerk registers the patient's identification data into the laboratory information system (HIS). Sampling and blood collection is done by the laboratory staff. All samples were collected in 2012 and there was no exclusion criterion.

After transporting specimens to the laboratory department, a laboratory technologist evaluates the samples for misidentification errors, inappropriate container, improper labeling, inadequacy of sample collection, and inadequacy of sample/anticoagulant ratio volume. Any probable errors are reported to the ordering department and request is sent for a new proper sample. Also, the errors are recorded in a roster and a copy is sent to the laboratory's quality control committee (the quality control committee comprises of a laboratory director, who is a professor of pathology; a supervisor, who is a graduate of laboratory sciences with at least 15 years of experience; and two laboratory technologists with at least 10 years of experience in quality control). The specimen preparation, including preparing suitable samples for analytic procedures like centrifugation, aliquoting, and sorting specimens are performed. Then, in a different laboratory section, tests are done. The test results data are entered to HIS both manually and automatically, depending on the test type. After monitoring and checking by a quality manager who is a technologist or a laboratory physician, the test results are allowed to be observed by medical staff for evaluation, interpretation, and appropriate action.

The physicians and nurses of hospital's departments are trained to be observant of test results and are asked to report any questionable result to the quality control committee of the laboratory. If any result is labeled clinically questionable, the quality manger checks the testing process and usually the specimen is retrieved or in some circumstances asks for a new sample. Errors are recorded by quality manager on a special roster and the date, time, source of the test, type and detail, and signature are recorded.

The reported case was discussed in the committee based on Clinical and Laboratory Standards Institute (CLSI) instruction [9] and classified in database. The quality control committee comprised the laboratory director, faculty member, QC manager as well as supervisor and sub-directors.

The laboratory error definition that has been incorporated in the International Organization for Standardization (ISO) Technical Report 22367 is defined as a defect occurring at any part of the laboratory cycle, from ordering test to reporting, interpreting, and reacting to results. [10] Since studies on laboratory errors are diverse and there is no general acceptance on definitions used especially on the types of errors, it is an arduous task to categorize variable types of errors. However, in this study it is tried to utilize more general and practical categorization as Da Rin used in his article. [11]

The laboratory testing process is divided into three phases:

1. Preanalytical Errors:

These errors are defined as the errors occurring from physician order to analytic phase, including test request, patient identification, collection, transportation, and preparation for analysis. For better classification and monitoring of preanalytical phase, it is subdivided to errors that occur outside the laboratory and errors that occur under the control of laboratory as Da Rin has mentioned in his study. [11] Outside Laboratory: inappropriate test request, order entry errors, misidentification of patient, inappropriate container, improperly labeled container, inadequate sample collection and transportation, specimen collected from infusion route, inadequate sample/anticoagulant volume ratio, and insufficient sample volume.

Within laboratory: sorting and routing errors, pour-off errors, and labeling errors.

2. Analytical Errors:

These errors occur during the test, and include equipment malfunction, sample mix-ups, interference, undetected failure in quality control, and procedure not followed.

3. Postanalytical Errors:

These errors happen after the test is conducted, and include failure in reporting, erroneous validation of analytical data, improper data entry, and excessive turnaround time.

Statistical analysis

The sum of errors was calculated. Their relative frequencies when compared with the total specimens were also calculated and presented as percentage. The statistical significance level accepted in this study was set at P value equal to or less than 0.05. The difference between relative frequencies of mistakes observed in the departments considered was tested by proportional Z test.


  Results Top


A total of 303,866 samples, 2,430,928 tests were received for analysis (230,938 samples, 1,847,504 tests were from inpatients and 72,928 samples, 583,424 tests were from outpatients). Total number of errors was 153,148 (6.3% - 62999.8 parts per million, ppm) (116392-62999.5 ppm for inpatients and 36756-63000 ppm for outpatients). Number and frequency of three main groups of errors in inpatients and outpatients are shown in [Table 1], and type and subtype of three main groups of errors in inpatients and outpatients are shown in [Table 2] and [Table 3], respectively.
Table 1: Number, Frequency and 95% CI of Errors in Preanalytical, Analytical, Postanalytical Phases

Click here to view
Table 2: Error and Sub-error Type Tests in Inpatients

Click here to view
Table 3: Error and Sub-error Type Tests in outpatients

Click here to view



  Discussion Top


Medical laboratory test results are instrumental in the decision-making by clinical doctors. Some studies have found that about 70% of clinicians' decisions are based on laboratory results. Therefore, the quality, accuracy, and precision of laboratory results are indispensable in clinical care.

According to a definition acknowledged by international organization for standardization, laboratory error is defined as "failure of planned action to be completed as intended, or use a wrong plan to achieve an aim, occurring at any part of the laboratory cycle, from ordering examinations to reporting results and appropriately interpreting and reacting to them." [10] The aim of this descriptive study is to evaluate type and frequencies of errors in three phases of laboratory tests to reduce preventable errors. The analysis of results of previous studies shows that despite heterogeneity in methods, the error distribution across the different phases of testing process appears to be similar with most errors occurring in the preanalytical phase. [11]

In one review article written by Howkins in 2012, [12] he mentioned that the proportion of errors associated with pre- and postanalytical phases of testing is 4-5 times higher than that seen in analytical phase with preanalytical phase representing over half of the errors in published studies. In another study, preanalytical factor consists of 46-68.2% of total errors while a high error rate 18.5-47% of total errors has been found in postanalytical phase. [13]

In a retrospective study performed by Plebani et al., [14] an Italian stat laboratory was assessed once in 1996 and then in 2006. They found out that despite a 34% reduction in error rate, the pattern of 62% preanalytical, 15% analytical, and 23% postanalytical phase errors remained basically unchanged. Analysis of results of this study shows that about 65.09% of errors occur in preanalytical phase, while about 23.2% and 11.68% occur across analytical and postanalytical phases, respectively.

According to different studies, there is a considerable difference between in- and outpatients with the 0.60% versus 0.039% for the two categories, respectively. [11] It seems that most of these differences are related to human factors including personal skills in venipuncture (drawing blood) and the sheer volume of laboratory tests carried out for inpatients. [15],[16] The results of this study also revealed that errors in hospitalized patients were more numerous than in outpatients. That might be due to the grave conditions of hospitalized patients and variety of staff involved in the total testing processes. For instance, the samples may be taken by nonexpert staffs. Ambiguities about the standard methods and appropriate transporting times for different laboratory tests are also of great importance in decreasing preventable errors. Therefore, nurses and other laboratory workers need to be more specialized in taking blood samples and other interventions. The samples need to be carried in the shortest possible time by experienced individuals. The vacutainers (glass or plastic tubes) could be prevented from being broken if they are carried in special vessels.

Although the importance of preanalytical phase has been acknowledged for many years, laboratories have often overlooked this area in their quality management programs, focusing instead on analytical quality and the associated activities within their direct control. [12],[17] In the analytical stage of this study, errors for inpatients and outpatients did not differ meaningfully and were mainly due to factors that could be prevented by an accurate and precise quality control procedure. Training special staff to conduct quality control or even offering a specialized university course is recommended for this purpose. In the postanalytical stage, the errors did not show any meaningful difference between these two groups and they resulted mainly from human errors in reporting procedures. To that effect, laboratories with high volume of work are recommended to be equipped with HIS system. In this way, the staff could be undercut and errors would be minimized. Postanalytical errors are indicative of the significance of thorough control on the results of the tests by the laboratory's chief technician before being confirmed by the laboratory director.

When compared with other studies, [9],[11],[15] this study shows more analytic errors, most likely because the aging equipment's state-of-the-art technology is not used for financial restrictions, or possibly, inadequate training of laboratory workers.

The limitations in this study were that registration of the errors during the evening, night-shifts, and holidays were not completely performed. More studies are recommended to be carried out to examine differences between day and night shifts separately and at multi centers. Hospital and laboratory directors can minimize errors and boost the quality of healthcare services based on the results of these studies.


  Conclusion Top


Since more than half of the laboratory errors occur during preanalytical phase, proper training and knowledge of the intervening factors that can influence laboratory results are essential to minimize laboratory errors. It should be reminded that all three stages of laboratory tests need to be under thorough monitoring to improve the quality of results.


  Acknowledgment Top


The authors thank all the staff at the laboratory who assisted and contributed to this study. The authors declare that there is no conflict of interests.

 
  References Top

1.Carraro P, Plebani M. Errors in a stat laboratory:types and frequencies 10 years later. Clin Chem 2007;53:1338-42.  Back to cited text no. 1
    
2.Plebani M, Carraro P. Mistakes in a stat laboratory: Types and frequency. Clin Chem 1997;43:1348-51.  Back to cited text no. 2
    
3.Hooijberg E, Leidinger E, Freeman KP. An error management system in a veterinary clinical laboratory. J Vet Diagn Invest 2012;24:458-68.  Back to cited text no. 3
    
4.Ibrahim F, Dosoo D, Kronmann KC, Ouedraogo I, Anyorigiya T, Abdul H, et al. Good clinical laboratory practices improved proficiency testing performance at clinical trials centers in Ghana and Burkina Faso. PLoS One 2012;7:e39098.   Back to cited text no. 4
    
5.Lanza LG, Stagi L. Non-parametric error distribution analysis from the laboratory calibration of various rainfall intensity gauges. Water Sci Technol 2012;65:1745-52.  Back to cited text no. 5
    
6.Aakre KM, Langlois MR, Watine J, Barth JH, Baum H, Collinson P, et al. Critical review of laboratory investigations in clinical practice guidelines: Proposals for the description of investigation. Clin Chem Lab Med 2013;51:1217-26.   Back to cited text no. 6
    
7.Chhillar N, Khurana S, Agarwal R, Singh NK. Effect of pre-analytical errors on quality of laboratory medicine at a neuropsychiatry institute in north India. Indian J Clin Biochem 2011;26:46-9.  Back to cited text no. 7
[PUBMED]    
8.Romero A, Cobos A, Gómez J, Muñoz M. Role of training activities for the reduction of pre-analytical errors in laboratory samples from primary care. Clin Chim Acta 2012;413:166-9.  Back to cited text no. 8
    
9.National Committee for Clinical Laboratory Standards. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. Approved Standard M7-A4. Wayne, PA: NCCLS; 1997.  Back to cited text no. 9
    
10.ISO/WD TS 22367. Medical laboratories-reduction of error through risk management and continual improvement. Geneva: International organization for standardization; 2008.  Back to cited text no. 10
    
11.Da Rin G. Pre-analytical workstations: A tool for reducing laboratory errors. Clinl Chim Acta 2009;404:68-74.  Back to cited text no. 11
    
12.Lippi G, Guidi GC, Mattiuzi C, Plebani M. Preanalyticalvariability: The dark side of the moon in laboratory testing. Clin Chem Lab Med 2006;44:358-65.  Back to cited text no. 12
    
13.Hawkins R. Managing the pre and postanalytical phases of the total testing process. Ann Lab Med 2012;32:5-16.  Back to cited text no. 13
[PUBMED]    
14.Plebani M. Errors in clinical laboratories or error in laboratory medicine?. Clin Chem Lab Med 2006;44:750-9.  Back to cited text no. 14
[PUBMED]    
15.Carraro P, Plebani M. Eroors in a stat laboratory: Types and frequencies 10 years later. Clin Chem 2007;53:1338-42.  Back to cited text no. 15
    
16.Bonini PA, Plebani M, Ceriotti F, Rubolli F. Errors in laboratory medicine. Clin Chem 2002;48:691-8.  Back to cited text no. 16
    
17.Karla J. Medical errors: Impact on clinical laboratories and other critical areas. Clin Biochem 2004;37:1052-62.  Back to cited text no. 17
    



 
 
    Tables

  [Table 1], [Table 2], [Table 3]


This article has been cited by
1 Short-term intra-individual variation of urinary biomarkers in dogs with stable chronic kidney disease
Hilla Chen, Lotem Maron, Gilad Segev
Journal of Veterinary Internal Medicine. 2023;
[Pubmed] | [DOI]
2 Complete Blood Count Collected Via Venipuncture Versus Peripherally Inserted Central Catheter in Hematological Patients
Elena Ponticelli, Marco Clari, Martina Volpes, Laura Valenti, Marco Angioni, Daniele Granero, Mattia Bert, Alessio Conti, Sara Campagna
Cancer Nursing. 2022; 45(1): E36
[Pubmed] | [DOI]
3 Taking a step back from testing: Preanalytical considerations in molecular infectious disease diagnostics
Stephanie Conrad, Anastasia Gant Kanegusuku, Steven E. Conklin
Clinical Biochemistry. 2022;
[Pubmed] | [DOI]
4 Assessment of Types and Frequency of Errors in Diagnostic Laboratories Among Selected Hospitals in East Wollega Zone, Oromia, Ethiopia
Edosa Kifle Tola, Yosef Tsegaye Dabi, Gemechu Tiruneh Dano
Pathology and Laboratory Medicine International. 2022; Volume 14: 1
[Pubmed] | [DOI]
5 The Forgotten (Invisible) Healthcare Heroes: Experiences of Canadian Medical Laboratory Employees Working During the Pandemic
Basem Gohar, Behdin Nowrouzi-Kia
Frontiers in Psychiatry. 2022; 13
[Pubmed] | [DOI]
6 Total Clinical Chemistry Laboratory Errors and Evaluation of the Analytical Quality Control Using Sigma Metric for Routine Clinical Chemistry Tests
Mulugeta Teshome,Abebaw Worede,Daniel Asmelash
Journal of Multidisciplinary Healthcare. 2021; Volume 14: 125
[Pubmed] | [DOI]
7 Analyse des risques appliquée au processus pré-analytique dćun laboratoire d’histocompatibilité
Imane Yakhlef,Sanae Ouadghiri,Ouafa Atouf,Malika Essakalli
Revue Francophone des Laboratoires. 2021; 2021(531): 74
[Pubmed] | [DOI]
8 False elevations in urinary metanephrines: under-recognised pitfall with 24-hour urinary volume collection
Terry Shin,Thanh Duc Hoang,Mary Thomas Plunkett,Mohamed K M Shakir
BMJ Case Reports. 2021; 14(2): e241147
[Pubmed] | [DOI]
9 Frequency and classification of errors in laboratory medicine at a veterinary teaching hospital in the United States
Kellie M. Whipple,Mary K. Leissinger,Sarah S. Beatty
Veterinary Clinical Pathology. 2020;
[Pubmed] | [DOI]
10 Day-to-day variation of the kidney proximal tubular injury markers urinary cystatin C, KIM1, and NGAL in patients with chronic kidney disease
Inga Soveri,Johanna Helmersson-Karlqvist,Bengt Fellström,Anders Larsson
Renal Failure. 2020; 42(1): 400
[Pubmed] | [DOI]
11 Effect of transportation and freeze-thaw procedure on hemostatic tests
Belkiz Öngen Ipek,Mustafa Erinç Sitar
Turkish Journal of Biochemistry. 2020; 0(0)
[Pubmed] | [DOI]
12 Estaciones automatizadas preanalíticas en el laboratorio de hemostasia. Estudio observacional descriptivo prospectivo, realizado en un hospital universitario de referencia entre el 15 de abril y 15 de julio de 2017
Luz Fernanda Sua,María Alejandra Amézquita,Daniel Esteban Hernández,Mercedes Alcalá-Flores,Carl Sebastian Leib,Marisol Aguirre-Rojas,Liliana Fernández-Trujillo
Iatreia. 2019; 32(3): 177
[Pubmed] | [DOI]
13 Analysis of preanalytical phase errors in a medical diagnostic laboratory
Alina Rak-Pasikowska,Dominika Goniwiecha,Iwona Bil-Lula
Diagnostyka Laboratoryjna. 2019; 55(2): 91
[Pubmed] | [DOI]
14 Barriers and Facilitators Impacting Reliability of the Electronic Health Record-Facilitated Total Testing Process
Daniel R. Murphy,Tyler Satterly,Deevakar Rogith,Dean F. Sittig,Hardeep Singh
International Journal of Medical Informatics. 2019;
[Pubmed] | [DOI]
15 Diagnostic Errors in Clinical Microbiology and Antimicrobial Resistance
Tulsi Chugh
Current Medicine Research and Practice. 2019;
[Pubmed] | [DOI]
16 Whatćs Old Is New Again: Laboratory Oversight of Point of Care Testing—Guidelines, Challenges, and Practical Strategies
Robert L. Sautter,Dawn M. Earnest,Diane C. Halstead
Clinical Microbiology Newsletter. 2018; 40(23): 191
[Pubmed] | [DOI]
17 Improvements in medical quality and patient safety through implementation of a case bundle management strategy in a large outpatient blood collection center
Shuzhen Zhao,Lujia He,Chenchen Feng,Xiaoli He
Medicine. 2018; 97(22): e10990
[Pubmed] | [DOI]
18 Nurses’ knowledge on phlebotomy in tertiary hospitals in China: a cross-sectional multicentric survey
Qian Cai,Yunxian Zhou,Dangan Yang
Biochemia Medica. 2018; 28(1)
[Pubmed] | [DOI]
19 Errors in the Hematology Laboratory at St. Paul’s Hospital Millennium Medical College, Addis Ababa, Ethiopia
Hirut Tadesse,Kassu Desta,Samuel Kinde,Fatuma Hassen,Addisu Gize
BMC Research Notes. 2018; 11(1)
[Pubmed] | [DOI]
20 ABOUT THE STANDARDIZATION AND QUALITY OF LABORATORY TESTS AIMED AT DIAGNOSTICS AND MONITORING OF TUBERCULOSIS CHEMOTHERAPY
D. V. V?KHRUSHEV?,I. ?. V?SILYEV?
TUBERCULOSIS AND LUNG DISEASES. 2018; 96(9): 57
[Pubmed] | [DOI]
21 Preanalitik süreçte kan alma hatalarinin incelenmesi
ÖZLEM AYDIN,Gülnihan GÖL,Füsun GÖNEN DEDE,Dilek AKIN
TURKISH JOURNAL of CLINICS and LABORATORY. 2017;
[Pubmed] | [DOI]
22 Prevalence of Pre-Analytical Errors in Clinical Chemistry Diagnostic Labs in Sulaimani City of Iraqi Kurdistan
Dereen Najat,Pal Bela Szecsi
PLOS ONE. 2017; 12(1): e0170211
[Pubmed] | [DOI]
23 Quantitative assessment of prevalence of pre-analytical variables and their effect on coagulation assay. Can intervention improve patient safety?
Ravi Bhushan,Arijit Sen
Medical Journal Armed Forces India. 2017;
[Pubmed] | [DOI]
24 Pre-Analytic Process Linked to Spuriously Elevated HIV Viral Loads: Improvement on an FDA-Approved Process
Gary W. Procop,Alan J. Taege,Colleen Starkey,Marisa Tungsiripat,Diane Warner,Jesse D. Schold,Belinda Yen-Lieberman
Diagnostic Microbiology and Infectious Disease. 2016;
[Pubmed] | [DOI]
25 Performance criteria of the post-analytical phase
Kenneth Sikaris
Clinical Chemistry and Laboratory Medicine (CCLM). 2015; 53(6)
[Pubmed] | [DOI]
26 Error rates during blood collection in emergency departments and outpatient clinics: Results of a prospective multicenter study
Giuseppe Lippi,Iole Caola,Gianfranco Cervellin,Bruno Milanesi,Margherita Morandini,Davide Giavarina
Clinica Chimica Acta. 2015; 445: 91
[Pubmed] | [DOI]
27 ISO 15189:2012 Management Requirements for Cytopathology Laboratory Information Systems
Abraham Pouliakis,Elena Athanasiadi,Efrossyni Karakitsou,Stavros Archondakis,Antonia Mourtzikou,Marilena Stamouli,Aris Spathis,Christine Kottaridi,Petros Karakitsos
International Journal of Reliable and Quality E-Healthcare. 2014; 3(3): 37
[Pubmed] | [DOI]
28 ISO 15189:2012 Technical Requirements for Cytopathology Laboratory Information Systems
Abraham Pouliakis,Niki Margari,Aris Spathis,Christine Kottaridi,Marilena Stamouli,Antonia Mourtzikou,Stavros Archondakis,Efrossyni Karakitsou,Elena Athanasiadi,Petros Karakitsos
International Journal of Reliable and Quality E-Healthcare. 2014; 3(3): 58
[Pubmed] | [DOI]
29 The journey toward quality and patient safety in laboratory medicine continues
Mario Plebani
North American Journal of Medical Sciences. 2014; 6(5): 229
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Materials and Me...
Results
Discussion
Conclusion
Acknowledgment
Introduction
References
Article Tables

 Article Access Statistics
    Viewed5089    
    Printed78    
    Emailed0    
    PDF Downloaded926    
    Comments [Add]    
    Cited by others 29    

Recommend this journal